As a pulsed UV laser, the MNL X00 is the ideal light source for bioreader systems and MALDI-TOF imaging. Compact, reliable and cost-effective, it delivers pulse energies of up to 110 µJ and pulse repetition rates of up to 150 Hz (200 Hz) on request.

  • Long lifetime
  • Inexpensive
  • For reader applications
  • Alternative to flash lamps
MNL X00 -UV Laser
Wavelength 337.1 nm
Typ. pulse energy @ number of shots depending on model 110@100 / 60@200 / 50@200 µJ @ Mio
Pulse power, typ. depending on model 20 to 37 kW
Energy stability ≤ 2 %
Focus stability < 15 µm
Current consumption 2.4 A

The Mini Nitrogen Laser MNL X00 is characterized by its long service life, low energy drop and high temporal precision. This is achieved by the following patented innovations:

  • Sealed laser tube in metal-ceramic technology
  • Direct switching semiconductor power switch

Up to 400 million laser pulses or maintenance-free operation for up to 4 years are guaranteed. An integrated laser controller enables extensive presettings and simple adaptation to different applications. The firmware offers the option of setting and controlling the laser functions and parameters via a PC interface. All trigger inputs and outputs are monitored by the trigger management system. The laser is supplied with an extra-low voltage of 24 V DC and is air-cooled. The scope of delivery includes a wide-range power supply unit (90 – 260 V, 50 – 60 Hz).

The MNL X00 is largely compatible with lasers of this class on the market and has the necessary certifications for all international markets: CE, ETL Report No. 05KFI005386 (UL, CSA, VDE, Semco), ROHS, FDA. Adaptation kits for installation in MALDI-TOF mass spectrometers or LIF systems from various manufacturers are available from LTB on request.

Features

Basic features

  • Pulse repetition rate: up to 60 Hz
  • Versions: Also available as a low-divergence version
  • Integrated control unit: Yes
  • Trigger management: Yes
  • Laser pulse counter, calibration, logbook function: Yes
  • Serial bus protocol, DLL interface: Yes
  • Interface converter RS 232 to optical: Yes

Options

  • Operation also possible without a PC
  • Laser energy can be varied by the customer
  • Integrable energy monitor
  • Integratable continuous beam attenuation unit up to 1 : 10,000
  • Optical sync. out (jitter < 200 ps)
  • Integrated fiber coupling
  • Laser can be equipped with a fiber (200 - 1,000 μm)
  • USB to RS 232 adapter

Basic specifications

Wavelength 337.1 nm
Spectral bandwidth 0.1 nm
Guaranteed pulse energy depending on model ≥ 60 to ≥ 110 µJ
Typ. pulse energy @ number of shots depending on model 110@100 / 60@200 / 50@200 µJ @ Mio
Pulse half-width FWHM, typ. 3 ns
Pulse power, typ. depending on model 20 to 37 kW
Energy stability ≤ 2 %
Beam dimensions, vertical x horizontal, typ. 3 x 4 mm
Divergence, vertical x horizontal 2 ≤ 3,5 x ≤ 3 mrad
Focus stability < 15 µm
Beam exit angle, vertical x horizontal, typ. 0,5 (± 0.2) x 0 ± 0.1 grad
Trigger In optical or electrical (TTL)

Electrical connection requirements

Voltage supply 24 V DC
Current consumption 2.4 A

Operating conditions

Operating temperature +15…+38 °C
Storage temperature -10…+60 °C
Max. relative humidity 85 %
Air pressure 750 – 1.300 mbar
Dimensions 335 x 95 x 95 mm
Weight 3.5 kg
Power supply dimensions (L x W x H) 180 x 80 x 50 mm
Weight of the power supply unit 0.6 kg

Models

UnitX06X12X15
Wavelengthnm337.1337.1337.1
Spectral bandwidthnm0.10.10.1
Pulse half-widthns333
Guaranteed number of shotsMio400120150
Guaranteed pulse energyµJ≥ 110≥ 70≥ 60
Pulse energyµJ110@10060@20050@200
Pulse power, typ.kW372320
Follow-up frequencyHz60120150
Energy stability SD%≤ 2≤ 2≤ 2
Beam dimensions, vertical x horizontal, typ.mm3 x 43 x 43 x 4
Divergence, vertical x horizontalmrad≤ 3,5 x ≤ 3≤ 3,5 x ≤ 3≤ 3,5 x ≤ 3
Focus stabilityµm< 15< 15< 15
Beam exit angle, vertical x horizontal, typ.deg0,5 (± 0.2) x 0 ± 0.10,5 (± 0.2) x 0 ± 0.10,5 (± 0.2) x 0 ± 0.1
Trigger Inoptical or electrical (TTL)optical or electrical (TTL)optical or electrical (TTL)
Jitter: ext. trigger - laser pulsens± 2,5± 2,5± 2,5
Pulse delay: ext. trigger - laser pulsens1.300 ± 10%1.300 ± 10%1.300 ± 10%
Sync Out (optional)
Jitter: elektr. Trigger – Laserpulsns< 0,2< 0,2< 0,2
Warm-up times< 20< 20< 20
Control systemautomode or software (DLL) via integrated controllerautomode or software (DLL) via integrated controllerautomode or software (DLL) via integrated controller
Guarantee / Warranty400 million pulses or 4 years120 million pulses or 2 years150 million pulses or 2 years
ZertifizierungenCE, ETL (ANSI/UL 61010-1, CAN/CSA C22.2#61010-1), FDA1CE, ETL (ANSI/UL 61010-1, CAN/CSA C22.2#61010-1), FDA1CE, ETL (ANSI/UL 61010-1, CAN/CSA C22.2#61010-1), FDA1
Laser classification3B / IIIb3B / IIIb3B / IIIb

MNL X00 -UV Laser

MNL 100

Compare product

Applications

  • OEM laser source
  • LIF
  • FRET
  • MALDI-TOF MS
  • MALDI Imaging
  • Ion traps MS
  • UV microscopy
  • µ-LIBS
  • Pumping dye lasers
  • Photoacoustics
  • Optoacoustics